Description
Technical data
| Parameter | Value | |
| Model | PC200 | |
| Dimensions (H x W x D) | 240 mm x 75 mm x 57 mm | |
| Weight | 570 g | |
| Light source | Laser class 1 (class 3R laser enclosed in a way to prevent tampering, 780 nm, 1.5-3 mW, classified according to DIN EN 60285-1 and the Technical Regulations of the Occupational Health and Safety Ordinance on Artificial Optical Radiation (TROS)) |
|
| PC interfaces | USB port | |
| Tripod thread | 1/4 inch – 20 UNC | |
| Storage conditions | -10 °C to +60 °C at 10 to 90 % RH (non-condensing) |
|
| Operating conditions | 0 °C to +50 °C at 10 to 90 % RH (non-condensing) |
|
| Display | 2.8 inch colour LCD display with background illumination, 320 x 240 pixels | |
| Functions | Maximum, minimum and average value display, measured value hold function, alarm function, language selection, °C/ °F switch, photo or video recording |
|
| Image format, resolution | JPEG, 640 x 480 pixels | |
| Video format, resolution | 3GP, 320 x 240 pixels | |
| Data storage | 5000 data records on the internal flash memory (optional memory expansion with a microSD card: up to 16 GB) | |
| Energy | ||
| Battery | Polymer LI-ION battery | |
| Operating time | approx. 4 hours of continuous operation | |
| Charging time | approx. 2 hours with an alternating current adapter | |
| Automatic switch-off | 3 minutes, 15 minutes or 60 minutes | |
| Automatic screen switch-off | 90 seconds, 2 minutes or 4 minutes | |
| Parameter | Value | |
| Particle counter | ||
| Channels (detectable particle sizes) | 0.3 μm, 0.5 μm, 1.0 μm, 2.5 μm, 5.0 μm, 10.0 μm |
|
| Flow rate | 2.83 l/min. (0.1 ft³/min.) (=> 0.99 l/21 s) controlled by internal pump | |
| Counter mode | Cumulative, differential, concentration | |
| Counting efficiency | 50 % at 0.3 μm; 100 % for particles > 0.45 μm (according to ISO 21501) |
|
| Coincidence loss | 5 %, 2 million particles per 28.3 litres | |
| Zero count | 1 count/5 minutes (according to JIS B9921) | |
| Start delay | 1 to 100 seconds | |
| Sample inlet | Isokinetic probe | |
| Calibration | Using monodisperse latex particles (PSL particles; compliant with NIST) | |
| Light source of the measuring cell | Laser class 1 (class 3R laser enclosed in a way to prevent tampering, 780 nm, 1.5-3 mW, classified according to DIN EN 60285-1 and the Technical Regulations of the Occupational Health and Safety Ordinance on Artificial Optical Radiation (TROS)) |
|
Professional particle measurement with full documentation and control
The ISO 21501-4 compliant PC200 is a versatile particle counter for applications ranging from cleanroom testing and filter efficiency checks to workplace monitoring and quality assurance. All six particle size channels are displayed simultaneously on the large 2.8-inch colour LCD, supported by a colour status indicator and automatic acoustic alarm for fast identification of critical concentrations. In addition to particle size and quantity fractions, the PC200 also measures air temperature, humidity, dew point and wet bulb temperature. The integrated digital camera enables photo and video documentation of the measuring environment. With a powerful data logger for up to 5,000 records, expandable by up to 8 GB via MicroSD card, and fast USB data transfer, the PC200 offers complete, professional measurement documentation and analysis.




Reviews