Description
Applications and usage
- Can I connect external Pt-100 or Pt-1000 temperature probes to HWg-STE plus?
No, external Pt100 and Pt1000 probes are connected using the Converter 2xPt100 1W-UNI. This converter cannot be used with Poseidon1 products or HWg-STE (HWg-STE PoE) and HWg-STE plus web thermometers. - Can I use the converter with only one Pt-1000 or Pt-100 probe?
Yes. The second probe will either read as -999.9 (indication of disconnected sensor) or turn off automatically. - Is there a difference between 2-wire and 4-wire connection of the external temperature probe?
Yes, 4-wire connection allows the most accurate measurement. The difference is noticeable especially with longer wires. - I have an external sensor connected but Poseidon2 shows weird values (5000 °C and so on). What is the problem?
Double-check the wiring and connection of the Pt-100 or Pt-1000 probe, according to the manual. Are the jumpers for unused terminals properly in place? - Can I connect any Pt-100 or Pt-1000 probe?
Yes. You can use a genuine HW group PT100 probe, or any other Pt100 or Pt1000 temperature probe.